1. Main technical indicators
1. Studio size: 1000D×1000W×1000H mm, the inner volume is about 1000L
2. Dimensions: about 3400D×1400W×2010H mm, excluding the controller, test hole and other protruding parts.
3. Temperature range: -70℃~+150℃
4. Temperature fluctuation: ≤±0.5℃, normal pressure, no load
5. Temperature deviation: ±2℃, normal pressure, no load
6. Temperature uniformity: ≤2℃, normal pressure, no load
7. Heating rate: +20℃→+150℃≤60min
8. Cooling rate: +20℃→-65℃≤60min
9. Humidity range: humidity 20%~98%RH (temperature +20℃~+85℃)
10. Humidity deviation: ≤+2 -3%RH (>75%RH), ≤±5%RH (≤75%RH), under normal pressure and no-load conditions.
11. Pressure range: normal pressure ~ 0.5kPa
12. Depressurization rate: normal pressure~1.0kPa≤30min
13. Pressure recovery rate: ≤10.0kPa/min
14. Pressure deviation: normal pressure~40kPa:≤±2kPa, 40KPa~4kPa:≤±5%kPa, below 4kPa:≤±0.1kPa
15. Wind speed: frequency conversion adjustment
16. Power: about 50kW
17. Noise: ≤75dB(A), measured at a distance of 1 meter from the front of the box and a height of 1.2 meters from the ground.
18. Weight: 1900Kg
2. Rapid decompression device (optional)
In order to meet the requirements of rapid depressurization. An independent rapid pressure reduction box is processed. The rapid pressure reduction box is composed of shell components, pressure bearing components, gate components, interfaces and moving frames. Before rapid pressure reduction, users need to connect external pipes.
1. Studio size: 400mm deep x 500mm wide x 600mm long; the inner wall material is made of 3.0 SUS304/2B, and a 5mm square tube is used as a pressure-bearing reinforcing rib;
2. External dimensions: 530mm deep × 700mm wide × 880mm long, the outer wall material is made of 1.2mm cold-rolled steel plate, and the surface is sprayed white (the same color as the box);
3. The top of the box is reserved for connecting the pressure sensor interface, and the control sensor interface is located on the rear side of the box, which is convenient for the quick pressure reduction device to go out.
4. For the convenience of moving the quick pressure reducing device. Four Qidong lifting casters are installed under the frame; the moving frame is welded and processed by ordinary section steel, and the surface is treated with plastic spraying;
5. Rapid depressurization process: First, the vacuum pump of the test equipment lowers the fast depressurization box to 75.2kPa. In order to improve the pumping speed of the test box to the fast depressurization device, the test box is first pumped to about 1kPa, and the connection test is started. The electric valve of the tank equipment and the fast pressure reduction device realizes the fast pressure reduction function. When the valve reaches 18.8kPa, the valve is closed, and the pressure in the fast pressure reduction box can be kept constant through the auxiliary pumping (intake valve).
Schematic
3. Product implementation standards
1. GB/T2423.1-2008 Test A: Low temperature test
2. GB/T2423.2-2008 Test B: Low Temperature Test
3. GB/T 2423.3-2006 Test Cab: Constant Damp Heat Test
4. GB/T 2423.4-2008 Test Db: Alternating Damp Heat Test
5. GB/T2423.21-2008 Test M: Low Pressure Test Method
6. GB/T2423.25-2008 Test Z/AM: low temperature/low pressure comprehensive test
7. GB/T2423.26-2008 Test Z/BM: high temperature/low pressure comprehensive test
8. General requirements of GJB150.1A-2009
9. GJB150.2A-2009 Low pressure (altitude) test
10. GJB150.3A-2009 high temperature test
11. GJB150.4A-2009 low temperature test
12. GJB150.6-86 temperature-altitude test
13. GJB150.19-86 temperature-humidity-altitude test
14. DO16F rapid decompression test
15. GB/T 10586-2006 Technical Conditions of Damp Heat Test Chamber
16. GB/T 10590-2006 Technical conditions of high and low temperature and low pressure test chamber
17. GB/T 10592-2008 Technical Standard of High and Low Temperature Test Chamber
18. GB/T 5170.1-2008 General Provisions for Inspection Methods of Environmental Test Equipment for Electrical and Electronic Industries
19. GB/T 5170.2-2008 Test method for environmental test equipment for electrical and electronic products – Damp heat test equipment
20. GB/T 5170.5-2008 Test method for environmental test equipment for electrical and electronic products – Damp heat test equipment
GB/T 5170.10-2008 Test method for environmental test equipment for electrical and electronic products – high and low temperature and low pressure test equipment